1. Support Center
  2. General Information

Efficacy

Patented and GSA approved, the iCleanse family of products effectively kills pathogens that can lead to infection in as little as 15 seconds.

iCleanse Swift & Swift Mini

Organism Tested ID Time (in seconds) Notes Log Reduction vs Control Efficacy Results Lab
MRSA ATCC 33591 15   >4.98 100.00% ResInnova
OC43 (Human Coronavirus)   15   >2.98 99.90% ResInnova

iCleanse Swift XL

Organism Tested ID Time (in seconds) Notes Log Reduction vs Control Efficacy Results Lab
MRSA Clinical Strain 30   >6.13 100.00% ResInnova
VRE Clinical Strain 30   >6 100.00% UCONN
C. Diff ATCC 70057 120 Toxinotype tcdA-, tcdB-
Ribotype 038
Binary toxin gene cdtB not
amplified by PCR
Nontoxigenic
1.2 – > 2.2 90% – 99.99% YNHH
OC43 (Human Coronavirus)   30   >3.5 100.00% ResInnova

iCleanse Swift 5x

Organism Tested ID Time (in seconds) Notes Log Reduction vs Control Efficacy Results Lab
Enterococcus faecium ATCC 51559 60 Multidrug-resistant (ampicillin,
ciprofoxacin, gentamicin, rifampin,
teicoplanin, and vancomycin) Presence
of VanAgene confirmed by PCR.
>4 99.90% YNHH
Staphylococcus aureus
subsp. Aureus (MRSA)
ATCC 43300 60 Resistant to methicillin, Resistant to
oxacillin.Absence of PVL gene
confirmed by PCR. MRSA
SCCmec: Type II
spa type: Ridom: t007
spa type Kreiswirth: WGKKKKAOM
pvl gene amplification: Negative
>4.5 100.00% YNHH
Pseudomonas aeruginosa ATCC 27853 60 Effect of salt concentration on gentamicin susceptibility >6 100.00% YNHH
Acinetobacter baumannii ATCC BAA747 60 Vitek #109234 [API 8310086, NSB 4221] >6 100.00% YNHH
Salmonella Clinical Strain 60   >5 99.90% YNHH
C.Diff ATCC 70057 120 Toxinotype tcdA-, tcdB-
Ribotype 038
Binary toxin gene cdtB not
amplified by PCR
Nontoxigenic
1 – > 3 90% – 99.90% YNHH

 

 

Results are based on tests of our devices in a laboratory setting. They are not meant to expand the claim of iCleanse’s ability to provide general-purpose disinfection to mobile devices.

  • The devices were inoculated and then processed in the iCleanse devices without pre-cleaning.
  • The process for C.Diff was by serial dilution (spores) & subsequent culture, not ATP.
  • Pre-cleaning will reduce the bio-burden before disinfecting and produce even better results.
  • The disinfection time can be changed based on your preferences by accessing the settings menu on your unit (refer to the user manual).
  • Remember, iCleanse provides an engineering control (automated, reproducible process) for disinfection, executing the same method in the same manner without relying on the variability of a user’s compliance with the protocol, affording the reduction or elimination of human error.