Efficacy

Patented and GSA approved, the iCleanse family of products effectively kills pathogens that can lead to infection in as little as 15 seconds.

iCleanse Swift & Swift Mini

Organism Tested ID Time (in seconds) Notes Log Reduction vs Control Efficacy Results Lab
MRSA ATCC 33591 15   >4.98 100.00% ResInnova
OC43 (Human Coronavirus)   15   >2.98 99.90% ResInnova

iCleanse Swift XL

Organism Tested ID Time (in seconds) Notes Log Reduction vs Control Efficacy Results Lab
MRSA Clinical Strain 30   >6.13 100.00% ResInnova
VRE Clinical Strain 30   >6 100.00% UCONN
C. Diff ATCC 70057 120 Toxinotype tcdA-, tcdB-
Ribotype 038
Binary toxin gene cdtB not
amplified by PCR
Nontoxigenic
1.2 – > 2.2 90% – 99.99% YNHH
OC43 (Human Coronavirus)   30   >3.5 100.00% ResInnova

iCleanse Swift 5x

Organism Tested ID Time (in seconds) Notes Log Reduction vs Control Efficacy Results Lab
Enterococcus faecium ATCC 51559 60 Multidrug-resistant (ampicillin,
ciprofoxacin, gentamicin, rifampin,
teicoplanin, and vancomycin) Presence
of VanAgene confirmed by PCR.
>4 99.90% YNHH
Staphylococcus aureus
subsp. Aureus (MRSA)
ATCC 43300 60 Resistant to methicillin, Resistant to
oxacillin.Absence of PVL gene
confirmed by PCR. MRSA
SCCmec: Type II
spa type: Ridom: t007
spa type Kreiswirth: WGKKKKAOM
pvl gene amplification: Negative
>4.5 100.00% YNHH
Pseudomonas aeruginosa ATCC 27853 60 Effect of salt concentration on gentamicin susceptibility >6 100.00% YNHH
Acinetobacter baumannii ATCC BAA747 60 Vitek #109234 [API 8310086, NSB 4221] >6 100.00% YNHH
Salmonella Clinical Strain 60   >5 99.90% YNHH
C.Diff ATCC 70057 120 Toxinotype tcdA-, tcdB-
Ribotype 038
Binary toxin gene cdtB not
amplified by PCR
Nontoxigenic
1 – > 3 90% – 99.90% YNHH

 

 

Results are based on tests of our devices in a laboratory setting and are not meant to expand the claim of iCleanse’s ability to provide general-purpose disinfection to mobile devices.

  • The devices were inoculated and then processed in the iCleanse without any pre-cleaning.
  • The process for C.Diff was by serial dilution (spores) & subsequent culture not ATP.
  • Pre-cleaning will reduce the bio-burden before disinfecting and produce even better results.
  • The disinfection time can be changed based on your preferences by accessing the settings menu on your unit (refer to user manual).
  • Remember, iCleanse provides an engineering control (automated, reproducible process) for disinfection, executing the same method in the same manner without relying on the variability of a user’s compliance to the protocol, affording the reduction or elimination of human error.